Point Load Strength

Supported Test Methods

Standard

Description

ISRM Part II:1985:6

Suggested Method for Determining Point Load Strength

Input

Lab-Rock | Point Load

To calculate the Is(50), you must first select the standard from the Test_Method field on the Point_Load table that the specimen is being tested against.

Enter the test data in the data entry fields in the lower table and clickSave, Ctrl+S or change table to initiate the calculation. The values will be calculated automatically and the result will be displayed in the calculated fields.

After a record is entered for the first time and saved, the first row will automatically copy down for fields where it might be useful.

The field Number_to_Exclude_from_Is50_Mean on Point_Load_Test (upper) table defines the number of highest and lowest values to exclude from point load strength test mean (eg. If Number_to_Exclude_from_Is50_Mean is 2, the two highest and the two lowest values will be excluded for the mean calculation). The excluded tests are shown in the Exclude_From_Mean check box.The net mean is written to the Point_Load_Test (upper) table.

Moisture contents for each specimen are recorded in the Moisture_Content table.

Output

  • Graphic Text Report | L R POINT LOAD STRENGTH SAMP
  • Histogram | A L R POINT LOAD INDEX
On this page